Manufacturing and Designing of Implantable 3D-Printed Medical Devices - An Overview of Regulatory Challenges
Amar S1, Balamuralidhara V2, T M Pramod Kumar3
1Regulatory Affairs Group, JSS Academy of Higher Education and Research, Sri Shivarathreeshwara Nagara, Mysore – 570015, Karnataka, India.
2Regulatory Affairs Group, Department of Pharmaceutics, JSS College of Pharmacy, JSS Academy of Higher Education and Research, Sri Shivarathreeshwara Nagara, Mysore – 570015, Karnataka, India.
3Principal, Department of Pharmaceutics, JSS College of Pharmacy, S.S.Nagara, Mysuru – 570015.
*Corresponding Author E-mail: amarbagirath.123@gmail.com, baligowda@jssuni.edu.in, pramodkumar@jssuni.edu.in
ABSTRACT:
From past several decades 3D printing technology has quickly came into the medical device sector and has been exploited by creative organization to produce products with distinctive composition, structure, and customizability. These distinguishing skills that 3D printing has created fresh regulatory difficulties. 3D printed devices customizability presents fresh Implications when developing an FDA market approval model for design control. The customizability and distinctive construction procedures of 3D printed medical devices present special difficulties in meeting the quality assurance of production regulatory standards. In order to guarantee a quality construction, reliable objects powder characteristics then Ideal parameters for printing such as construction orientation and laser energy should discussed later conveyed to FDA. Also addressed are post-printing factors that are special to three dimensional printed devices, like sterilization cleaning, and finishing. It will lead a controlled output and continuous device production by Use of the same quality control and design policies used to conventional built-up techniques with three-dimensional printing. The advancement of 3D printing technology has produced innovative medical devices with a new structure and constituents. This research shows the distinctive regulatory problems facing USFDA consideration to the device plan growth, production, sterilization and biocompatibility.
KEYWORDS: FDA, Quality, Biocompatibility, innovative, customizability.
INTRODUCTION:
Investigational Device Exemption (IDE) is generally used for class three medical device, which allows device to be used solely in a clinical trial regulated by the FDA to obtain vital efficacy and safety information earlier to market implementation. The Humanitarian Use Device (HUD) route enables candidate to create a device without an efficacy information necessity, the route is considerably expedited, with the applicant being certain a choice within 45 calendar days. In order to get a Humanitarian description, the creator must give the FDA with following paperwork stating the frequency of a disease that has to impact less than 4,500 patients per year in the US. in addition, the creator must consist of a complete explanation of the device and its proposed use, with scientific justification behind the device. Device is eligible for marketing through the Humanitarian Device Exemption (HDE) pathway once the applicant has approved the HUD description and the IDE requirements have been met., An implementation analogous to the PMA in that safety information given, but exempted in particular from the provision of information on efficiency. Three dimensional also known to be additive manufacturing or rapid prototyping, it is a development Whereby a three-dimensional product model is electronically cut into single layers and the material is layer-wise deposited until the final product is constructed.1
OBJECTIVE:
Overview of distinctive regulatory problems in the creation, manufacture, biocompatibility and sterilization of devices with regards to USFDA consideration.
DISCUSSION:
Summary of three-dimensional printing of medical devices.
· Medical device design is developed to computer-aided software (CAD) user needs and exported as a stereo lithography (STL) file.
· Device. STL is layer-wise "sliced" (left) and laid out using 3D printer software on the virtual build platform (right)
· Construction is exported to a 3D printer and assembled by depositing "slice-by-slice" material until to get the final model. S. D.1
1. Pre-Printing Consideration:
· We searched and analysed the inner FDA medical device database using AM-relevant keywords to monitor trends in using this technology.
· Many AM technologies used to define each and a varied lexicon. Furthermore, the sort of production is not a necessary element of the 510(k) overviews of the public.
· Search findings for applicability have been manually checked and misidentified entries filtered. Limited to keywords results explained, identified to ensure data confidentiality.
· The objective of this study was to provide for the latest history of this evolving technology helpful picture that kinds of unfurnished Additive manufactured instruments and additive manufactured procedures generate equipment.2
· Conditions that time frarne work chosen and dataset does not therefore constitute full catalogue Additive Manufactured devices that have been cleared by FDA.3
· More than 90 medical devices inward 510 (k) clearance within the limitations of the search described.
Figure 1: Various Categories involved in manufacture of 3D Printing implantable medical device (2)
Table 1: Common regulatory pathways of the FDA for the regulation of additive medicinal Products (4)
|
Pathway |
Description |
|
Premarket notification |
Application used for marketing authorization by indicating the device which is labelled is having slightest as effective as a safe, i.e. significantly equal officially marked product it is not applicable to premarket authorization |
|
Premarket approval (PMA) |
Application used to marketing permit Class three device that presents a substantial danger of disease or Injury or devices discovered not significantly equal to the 510(K) method and showing device's efficacy and safety. |
|
New drug Application (NDA) |
Application to the drug which is sale and marketing in the US showing it is secure efficient also |
|
Biologic license application |
Application to the authorization in which bring a product prepared biologically to the interstate trade and ensures it is efficient and secure. |
|
Master file |
Compliance Food and drug administration to give private comprehensive data on equipment, procedures, the production details on drug, biologic, device. A file may mentioned through statutory file filed to medical product authorization or permission. |
Manufacture of 3d Printing Implantable Medical Device in Various Categories:
· Most of the FDA-cleaned additive produced instruments discovered in this search were produced utilizing powder bed fusion methods (83%), including selective laser sintering (66%), electron beam melting (25%) and- other laser-based technologies (9%).
· These techniques can generate complex geometries with comparatively excellent spatial resolution. In addition, a range of conventional metallic and polymeric materials are suitable for these printing procedures to meet the elevated temperatures needed. In some other technologies fewer frequently used in Approved devices which is survey in the search consists stereo lithography (12%), extrusion (3%), and inkjet (2%).
· Clinical application analysis showed 53% unfurnished instruments recognized in these outcomes belongs to orthopaedic implants and 34% surgical guides. For dental (6%) and craniofacial (7%) apps, the remaining instruments were used. Spine (11 %) and knee (6%) are other orthopaedic applications.
Material selection depended heavily on implementation and planned use. By using materials which is used in conventional instruments most three-dimensional printed metal instruments produced.
Selection of materials strongly based on execution and scheduled use. Most of the three-dimensional printed tools of metal manufactured by means of the materials which is similar as standard tools, including alloys like. Depending that imaging medical, design files digitally these polymer systems were predominantly patient matched.
It is essential to note that in the context of a medical item and its planned use, the FDA evaluates a material which is not cleared products to general medical use alone. AM sector is growing quickly, with AM, specifically powder bed fusion techniques, being an early adopter of the orthopaedic device sector. Many of the first dental machines pro ducts to be cleaned were in the region. Most approved Additive manufacturing systems integrate permeable constructions and are specific to patient, taking advantage in which benefits from Additive manufacturing deliver through traditionally production methods.3,5
Design control:
Design control is a device construction scheme that encourages the detection of possible design faults, various design ideas, and design effectiveness verification and validation through frequent design review.
· Implantable devices of Class III typically involve before going to clinical and preclinical and usage of animal testing to confirm the efficiency of the device and to create a profile of safety. validate the TBS criteria for clinical results.
· The FDA not openly articulated FEA's regulatory consent as a substitute for standard device design validation. However, FEA can be a helpful mechanism for evaluating the efficiency of 3D-printed systems in device design.6,7
2. Printing Consideration:
Raw materials:
FDA Quality System (QS) Regulation/Medical Device Good Manufacturing Practices for industry manufactured devices should be implemented throughout the manufacturing process of 3D-printed implantable devices, beginning with the substrate used for manufacture
· Although many of the quality assurance (QA) processes for three dimensional printed devices are like those used in industry for decades, a unique manufacturing process permitted by three-dimensional printing is the "recycling" of manufacturing substrates from one construction to the next.8
· It remains to be determined whether the FDA will accept routine material recycling for 3D-printed implantable devices. In order to better characterize these hazards, further research is needed to identify the potential for material contamination and further shelf-life a sting of three dimensional printing biomaterials, and a thorough survey of the selected raw materials will be necessary for regulatory endorsement of three dimensional printed implants.9
Consideration by technically:
· Number of 3D printing parameters have been a shown to have a significant effect on the final produced device's physical features, such as laser beam power density, scanning speed, deposition velocity and moisture within the build setting. Therefore, quality measures are necessary and necessary to document for regulatory purposes to ensure consistency between builds.
· The FDA will require all TBS printing parameters and procedures to be included in the IDE application to reduce the risk of accuracy build. We have created a set variety of TBS construction parameters to guarantee consistency of PCL laser sintering based on our group's previous job. It includes 5-Watt laser power, 51- 55°C bed temperature, 1100- 1600mm/ sec laser scanning velocity, and 0.14-0.19mm scanning spacing.9
Post manufactures quality assurance:
Standard guidance for 3D-printe devices post-manufacture QA has not yet been released. We use calliper readings of wall thickness, inner diameter and length for the TBS and presently implement micro-CT procedures to evaluate the geometry and density of the device. The FDA considered these procedures to be adequate for the IDE application.10
3. Post Printing Consideration:
Cleaning and finishing:
· It will probably be necessary to validate that finishing procedures do not change the general structure or mechanical. Characteristics of the produced device beyond design input specifications but have not been stated explicitly by the FDA to date.
· After cleaning, all systems were mechanically screened to guarantee that the cleaning method did not substantially change the functional efficiency of the device, and this was considered suitable for incorporation into the IDE implementation. We do not use any finishing methods and therefore did not need to validate the machine after completing procedures.11
Biocompatibility
· The International Standards Organization (ISO) has released a collection of standards to assess the biocompatibility of materials used in the production of medical instruments, jointly referred to as ISO 10993
· These requirements have been enacted by the FDA and are being enforced for Class III instruments. Ultimately, the device structure, material composition and planned use determine which elements of ISO 1 0993 are required for regulatory authorization
· Therefore, raw material testing may be insufficient. Compounds throughout the production phase (such as supporting components, moving agents or bonding materials) may also have to be evaluated for abolishment during washing procedures or for compliance to ISO 10993 if they're still present in either the finished produced device.
· Specifically asked by the Food and drug administration to describe the mechanical output of the TBS over 9 months, accelerated in vitro degradation testing, where units were placed in an incubator in a basic solution.12,13
Sterilization:
Sterilization is essential to minimize the risk of implantable medical devices infection. Due to the various equipment used in 3D printing, it is possible to use a range of sterilization procedures.14
· Three dimensional printed implants involve the same verification of SAL like any other medical implant, independent of the sterilisation method, and the findings must be submitted to the FDA as part of the commercial application.
· In a comparable model to degradation testing, the FDA asked that sterility validation be carried out on limited TBS design factors. In addition, the FDA asked minimum criteria to show device security, which included in vitro cyto-toxicity research underneath the criteria of ISO 10993-5 for the TBS. 15
The Food and drug administration workshop and records concentrate mainly on medical devices, including some main factors for assessing additive produced devices, such as the impact of construction orientation and place on the efficiency of final equipment Along with technical and physical characteristics, method verification of Additive manufacturing systems to guarantee consistency among printing work, sterilization and withdrawal of remaining components, and product characteristics description pre and post printing.3
Every Food and drug administration Centre have programs to assist inventors and developers of medical products and the regulator has developed an agency-wide inner working group to promote intra-agency interaction and external involvement with partnerships and consortia in research and innovation.3
Adverse Events for Additive Manufactured Devices:
We performed post-market assessment of the approved devices recognized in the past chapter used the Manufacturer and user Facility Device Perspective database to better comprehend future safety and technical issues related to three dimensional printed medical devices.
During 2018 we gathered and published reports of medical devices (MDRs) recorded in MAUDE, which was the latest year at the moment of evaluation with full MDR data. In the study outlined above in this manuscript, we discovered 1260 reports connected with approved Additive manufacturing equipment.
These MDRs were tested to collate information on adverse outcomes (AEs) related to product-related additive produced devices, utilize errors, as well as therapy errors.
Since investigating of MDRs by producers, importers and device users is mandatory, but reporting by device users such as health care providers, patients and consumers is compulsory, and since the search included only systems found in the prior summary review, the effectiveness of the MAUDE database analysis was restricted. This evaluation may assist to define statistical parameters of device error or particular Adverse events for additive produced systems, given constraints.
Figure 1: Statistical representation of Adverse Event factors in 3D Printing implantable medical device2
Inappropriate size or incompatibility with many other device parts, device collision or shavings, insufficient or wrong system sizes for the patient, implant wear and use mistake (wrong device implanted) were the main factors for the item-related adverse events in priority order.
Most of these AEs are compatible through the use of traditionally produced instruments observed with AEs. It is hard to assume whether any of these AEs are directly linked to the production method or workflow due to the research scope and restricted information comprehensiveness.
The present study shows that although only a tiny proportion (9.3 percent) of the general Adverse events are item-related, the difficulty of Additive manufacturing instruments may give unique attention to extra technological and design factors required to avoid item faults.
With the exception of inventory instruments manufactured in one of many dimensions and widely evaluated prior to manufacturing, patient-matched devices [see FDA Custom Device Exemption Guidelines for more data may integrate any mix of shapes within a clear set of size parameters such as, angle, length, density etc. This is referred to as the design envelope.
Depending on this evaluation, inspecting the worst-case design in the package and evaluating the anatomical corresponding procedure functionality may be essential to avoid system error.3,16
CONCLUSION:
The advancement of three-dimensional manufacturing technology has produced unique medical equipment with a new structure and constituents. The capacity to quickly change the exact size of such appliances to satisfy particular patient requirements enables "custom" devices to be produced quickly, but creates unique regulatory difficulties for product design, production, biocompatibility and sterilisation. Regulatory requirements for the design and production of three-dimensional printed medical equipment Eventually, the growth and development of three-dimensional printing in the biomedical sector will depend on both the continuing development and development of legislation and technology. The application with the same quality and performance control strategies used in conventional three-dimension printing methods of production will result in a controlled output and consistent device manufacturing. Furthermore, considering the "custom" design of several three-dimensional printed implants, additional guidance is needed to determine which quality control steps are required of each device iteration.
ABBREVIATIONS:
IDE - Investigational Device Exemption
HUD - Humanitarian Use Device
CAD - Computer-aided software
AM - Additive manufacturing
QS - Quality System
MDR - Medical device regulation
ACKNOWLEDGEMENT:
The authors thank JSS Academy of Higher Education and Research and JSS College of Pharmacy, Mysuru for providing the necessary facilities for carrying out this work.
REFERENCES:
1. S. D. Mankar, Chaitrali Kale, Jangam Kanchan. 3D Printing Technology- A Computer Aided Design- A Review. Research J. Science and Tech. 2019; 11(3): 217-224.
2. Laura M. Ricles1, 2, James C. Coburn3 MDP and SSO. Regulating 3D-printed medical products [Internet]. Science Translational Medicine : Available from: https://stm.sciencemag.org/content/10/461/eaan6521.full
3. Review R of MD in RCAC, G.V.S.S.N. Jyothi, MPharm, M.P. Venkatesh, MPharm, PhD, T.M. Pramod Kumar, MPharm P. Regulations of Medical Devices in Regulated Countries: A Comparative Review [Internet]. Sage Journals; Available from: https://journals.sagepub.com/doi/abs/10.1177/2168479013492735
4. Kamalraj Raji, Vijey Aanandhi M. Regulatory Filing in Us and Eu: A Comparative View. Research J. Pharm. and Tech. 2017; 10(1): 286-292.
5. Harsha. L, Anand S. Literature Review on "Peek" Dental Implants. Research J. Pharm. and Tech 2016; 9(10): 1797-1801.
6. Green‡ SJHLFJMJPBWPEE. Integrating Image-Based Design and 3D Biomaterial Printing to create patient Specific Devices within a Design Control Framework for Clinical Translation [Internet]. ACS Publications; Available from: https://pubs.acs.org/doi/10.1021/acsbiomaterials.6b00332
7. Dubey R.D., Chandraker G., Sahu P.K., Paroha S., Sahu D.K., Verma S., Daharwal S.J., Prasad Reddy S.L.N. Computer Aided Drug Design: A Review. Research J. Engineering and Tech. 2011 2(3): 104-108.
8. Kovalyov Volodymyr, Chetvernia Serhii, Krechun Anastasiia, Mykhailenko Оlga. Establishment of the Main Technological Parameters of Iris raw material. Research J. Pharm. and Tech. 2019; 12(7): 3359-3364
9. Morrison RJ, Kashlan KN, Flanangan CL, Wright JK, Green GE, Hollister SJ, et al. Regulatory Considerations in the Design and Manufacturing of Implantable 3D-Printed Medical Devices. Clin Transl Sci. 2015; 8(5): 594–600.
10. Bhupendra Shrestha, Hema Basnett, P Mohan Raj, Sita Sharan Patel, Mrinmay Das, Neelesh Kumar Verma. Process Analytical Technology: A Quality Assurance Tool. Research J. Pharm. and Tech. 2009; 2(2): 225-227
11. Kamal Hossain, Kamrun Nahar, Ehsanul Hoque Mazumder, Tony Gestier, Tanvir Ahmed Khan, Kaiser Hamid. Development of a Cleaning Validation Protocol for an Odd Case Scenario and Determination of Methoprene residues in a Pharmaceutical Manufacturing Equipment Surfaces by using a Validated UFLC Method. Research J. Pharm. and Tech 2017; 10(11): 3789-3794.
12. Timothy M. Janetos, BSa, Comeron W. Ghobadi, MDb, Shuai Xu, MD, MSc (Lond)c, Jessica R. Walter Md. Overview of high-risk medical device recalls in obstetrics and gynecology from 2002 through 2016: implications for device safety [Internet]. American journal of Obstetrics and Gynecology; Available from: https://www.ajog.org/article/S0002-9378(17)30459-3/fulltext?rss=yes
13. Aslam Hossain, Sanjay Roy, Partha Sarathi Guin. The Importance of Advance Biomaterials in Modern Technology: A Review. Asian J. Research Chem. 2017; 10(4): 441-453.
14. Roberts CMSA. 3D and 4D Printing in Biomedical Applications: Process Engineering and Additive Manufacturing [Internet]. Maniruzzaman M, editor. Wiley Online Library, Jss Dental college and Hospital Library; Available from: https://onlinelibrary.wiley.com/doi/book/10.1002/9783527813704
15. Sowjanyaa J., Chandana C.S. Clinical Practice Guidelines on Sterilization of Endodontic Instruments. Research J. Pharm. and Tech. 2018; 11(3): 1039-1040.
16. Nukalapati Mishma Jaasu, Raju Kamaraj, R. Seetharaman. Med DRA (Medical Dictionary for Regulatory Activities). Research J. Pharm. and Tech 2018; 11(10): 4751-4754.
Received on 27.02.2020 Modified on 11.04.2020
Accepted on 11.05.2020 © RJPT All right reserved
Research J. Pharm. and Tech 2021; 14(3):1785-1790.
DOI: 10.5958/0974-360X.2021.00317.6